These models are not available in the US. Guidant CONTAK RENEWAL 4 RF/HE (model H230, H235) Guidant CONTAK RENEWAL 4 RF

FDA device recall Z-1050-2007 · posted 2007-07-11 · Terminated

Recall details

Recalling firm
Boston Scientific CRM Corp
Reason for recall
Boston Scientific has recently identified low-voltage capacitors from a former supplier that may be subject to degradation at a low rate of occurrence. These capacitors may cause accelerated battery depletion and may reduce the time between elective replacement indicator (ERI) and end of life (EOL) to less than 3 months.
Action taken
An Urgent Medical Devivce Information letter dated April 05, 2007 was sent to implanting and following physicians. The letter describes the issue, identifies patients with affected devices and provides recommendations.
Root cause
Other
Status
Terminated
Product code
NIK
Quantity affected
1,217
Distribution
Worldwide, including USA, Argentina, Aruba, Australia, Austria, Barbados, Belgium, Bermuda, Brazil, Canada, Chile, Costa Rica, Cyprus, Czech Republic, Czechoslovakia, Denmark, Dominican Republic, Ecua
Date initiated
2007-03-23
Date posted
2007-07-11
Date terminated
2008-03-02
Location
Saint Paul, MN