These models are not available in the US. Guidant CONTAK RENEWAL 4 (models H190, H195). Cardiac Resynchronization Therap

FDA device recall Z-1289-06 · posted 2006-07-27 · Terminated

Recall details

Recalling firm
Boston Scientific CRM Corp
Reason for recall
Guidant has identified a pattern of premature battery depletion in devices manufactured with a single lot of capacitors from a single supplier. If function of this low voltage capacitor is compromised, the battery may deplete prematurely.
Action taken
An Urgent Medical Device Safety Information & Corrective Action letter, dated May 12, 2006, was sent to physicians. The letter gave a description of the issue, clinical implications, rate of occurrence, devices affected and recommendation. Physicians are asked to follow-up with affected patients as soon as possible because ERI to EOL time may be shortened. Physicians are asked to contact Guidant for instructions on performing a baseline 'save to disk' and provide this information to Guidant. Guidant will be able to analyze the device memory data and estimate remaining longevity and provide individualized follow-up and replacement guidelines. Subsequent follow-up visits should continue to monitor for evidence of premature battery depletion.
Root cause
Other
Status
Terminated
Product code
NIK
Quantity affected
18
Distribution
Nationwide including Puerto Rico. Worldwide, including Argentina, Belgium, Canada, Denmark, France, Germany, Greece, Ireland, Italy, Netherlands, New Zealand, Portugal, Spain, Syrian Arab Republic, Un
Date initiated
2006-05-15
Date posted
2006-07-27
Date terminated
2008-03-02
Location
Saint Paul, MN