Atlas FDA

ThermoSuit Manufactured for Life Recovery Systems. HD, LLC, 170 Kinnelon Rd. Kinnelon, NJ 07405 USA Tel.: (973) 283-2800

FDA device recall Z-0456-2012 · posted 2012-01-11 · Terminated

Recall details

Recalling firm
Life Recovery Systems HD, LLC
Reason for recall
ThermoSuit System Model TSP-50 Pump/Controller may have corrosion on the temperature circuit board which may result in incorrect patient temperature reading.
Action taken
Life Recovery Systems sent recall letters/return response forms dated 10/24/2011 to all consignees. Life Recovery Systems has expanded their recall to include additional units and to provide additional service to products out in the field. A new recall letter was sent to consignees on 4/27/2012.
Root cause
Process control
Status
Terminated
Product code
NZE
Quantity affected
28 units
Distribution
Nationwide Distribution including NY, NC, LA, NJ, AL. PA TX, KY, MO, MA, and TN
Date initiated
2011-10-21
Date posted
2012-01-11
Date terminated
2013-01-24
Location
Waldwick, NJ

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
THERMOSUIT SYSTEM (K061023)Life Recovery Systems Hd, LLC2006-09-20