THERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. Used for catheter-based cardiac electroph
FDA device recall Z-0169-2023 · posted 2022-11-02 · Open, Classified
Recall details
- Recalling firm
- Johnson & Johnson Surgical Vision, Inc.
- Reason for recall
- Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation causing catheter tip temperature increase, limiting the ability to deliver ablation, or 3) particles flushed through the catheter irrigation may have thrombogenic/toxic effect.
- Action taken
- On 09/07/2022, recall notices were mailed and emailed to customers requesting the following: 1) Ensure that anyone in your facility who needs to be aware of this notification reads the notice carefully. 2) If any of the affected products have been forwarded to another facility, contact that facility, and provide a copy of this letter to the relevant personnel. 3) Discard/Destroy the device as per your standard disposal process, ensuring it cannot be used. 4) Complete and return the business response form. Questions can be directed to the recalling firm representative. The following email address was provided: OneMD-Field-Actions@its.jnj.com
- Root cause
- Process control
- Status
- Open, Classified
- Product code
- OAD
- Quantity affected
- 91
- Distribution
- Worldwide - US Nationwide distribution in the states of NC, TN, VA, OH, MO, MI, GA, OR, MO, SD, CO, OH, FL, NJ, TX, IL, WI and the countries of CROATIA, GERMANY, HUNGARY.
- Date initiated
- 2022-09-07
- Date posted
- 2022-11-02
- Location
- Irvine, CA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.