Atlas FDA

THERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. Used for catheter-based cardiac electroph

FDA device recall Z-0169-2023 · posted 2022-11-02 · Open, Classified

Recall details

Recalling firm
Johnson & Johnson Surgical Vision, Inc.
Reason for recall
Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation causing catheter tip temperature increase, limiting the ability to deliver ablation, or 3) particles flushed through the catheter irrigation may have thrombogenic/toxic effect.
Action taken
On 09/07/2022, recall notices were mailed and emailed to customers requesting the following: 1) Ensure that anyone in your facility who needs to be aware of this notification reads the notice carefully. 2) If any of the affected products have been forwarded to another facility, contact that facility, and provide a copy of this letter to the relevant personnel. 3) Discard/Destroy the device as per your standard disposal process, ensuring it cannot be used. 4) Complete and return the business response form. Questions can be directed to the recalling firm representative. The following email address was provided: OneMD-Field-Actions@its.jnj.com
Root cause
Process control
Status
Open, Classified
Product code
OAD
Quantity affected
91
Distribution
Worldwide - US Nationwide distribution in the states of NC, TN, VA, OH, MO, MI, GA, OR, MO, SD, CO, OH, FL, NJ, TX, IL, WI and the countries of CROATIA, GERMANY, HUNGARY.
Date initiated
2022-09-07
Date posted
2022-11-02
Location
Irvine, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.