Thermo Scientific HM 560 Cold Disinfection Cryostat, Microm International GmbH, Waldorf, Germany Tissue processing equip

FDA device recall Z-1735-2011 · posted 2011-03-21 · Terminated

Recall details

Recalling firm
Thermo Fisher Scientific
Reason for recall
Through internal quality review, firm determined the need to conduct a more comprehensive system validation to support the claim of complete bacterial, fungal and viral disinfection efficacy.
Action taken
ThermoFisher sent Advisory Notice letters beginning on September 28, 2010, to all affected customers. The letters stated that the firm needed to conduct more comprehensive system validation data on the Cold-Disinfection sytem option available on the HM550-D cryostat to support the claim of complete bacterial, fungal and viral disinfection efficacy. The letter recommended that the Cold Disinfection option on the HM550-D not be used as the sole disinfection method on the HM550-D until the validation work is completed. The firm provided addtional safety protocols pertaining to the cryostat disinfection. Customers were instructed to complete and fax the Confirmation of Receipt Form to Technical Support, COLD-D Advisory Notice at (269) 372-2436.
Root cause
Device Design
Status
Terminated
Product code
IDP
Quantity affected
100 units
Distribution
Worldwide Distribution - USA including the states of LA, OH, FL, MD, PA, TX, MO, WI, KS, AL, CA, VA, MS, IL, NE, MA, and the countries of Australia, Belgium, China, Czech Republic, Denmark, Estonia, F
Date initiated
2010-09-23
Date posted
2011-03-21
Date terminated
2012-04-02
Location
Kalamazoo, MI