The Verigene Enteric Pathogens Nucleic Acid Test (EP) is a multiplexed, qualitative test for simultaneous detection and
FDA device recall Z-2685-2014 · posted 2014-09-23 · Terminated
Recall details
- Recalling firm
- Nanosphere, Inc.
- Reason for recall
- Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures performed on the Processor SP caused by Tips in Tip Holder Assemblies for the following products and lots of Extraction Trays.
- Action taken
- Nanosphere sent a Field Notification Letter dated October 10, 2013, to all affected customers. The letter identified the product, the problem and the action to be taken by the customer. Instructions were included in the letter for customers to determine whether or not the extraction tray lots should be returned. If the extraction tray lots are to be returned, customers were to complete and return the attached form and replacement tip holder assemblies will be supplied at no cost. Customers could then dispose of the recalled unused tip holder assemblies. Customers with questions were instructed to contact their Nanosphere technical support representative. For questions regarding this recall call 1-888-837-4436, ext 3.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- PCH
- Quantity affected
- 220 units
- Distribution
- Worldwide Distribution - USA including AZ, CA, FL, GA, IA, IL, IN, KY, MA, MD, MN, MO, MT, NC, NE, NJ, NY, OH, OR, PA, TX, UT, VA, WA, WI, OH and Internationally to Bulgaria, Germany, Italy, Japan, So
- Date initiated
- 2013-10-10
- Date posted
- 2014-09-23
- Date terminated
- 2015-06-16
- Location
- Northbrook, IL