Atlas FDA

The Starion Instruments Thermal Ligating Shears 35C Model 102-138D and 23C Model numbers 102-137D allow the surgeon to p

FDA device recall Z-2239-2010 · posted 2010-08-19 · Terminated

Recall details

Recalling firm
Starion Instruments
Reason for recall
The Starion Instruments Thermal Ligating Shears (STLS) 35C and the Starion Thermal Ligating Shears (STLS) 23C may turn on and remain on without the finger button being depressed. The devices may also work intermittently, incorrectly indicating the device is 'on".
Action taken
Starion distributed a "STLS Device Recall - Important Medical Device Information" letter dated March 9, 2010 to consignees. The letter identified the affected model and lot numbers. Consignees were asked to complete a inventory status form and fax to Starion. the letter included a shipping labe with a pre-assigned return authorization number. Starion will ship replacement after receipt of the inventory status form and returned product..
Root cause
Device Design
Status
Terminated
Product code
GEI
Quantity affected
228 units
Distribution
Worldwide Distribution: USA states of CA, OH, MI, KS and MO and countries of Italy, China, Kuwait, and Germany.
Date initiated
2010-03-09
Date posted
2010-08-19
Date terminated
2010-11-27
Location
Sunnyvale, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
STARION INSTRUMENTS THERMAL LITIGATING SHEARS (K062257)Starion Instruments2006-10-10