The St. Jude Medical Livewire TC" Ablation Catheter, 8mm Tip w/Thermocouple & Thermistor REF 402196, Length 115 cm, Lot

FDA device recall Z-2313-2008 · posted 2008-09-20 · Terminated

Recall details

Recalling firm
St. Jude Medical Cardiovascular Division
Reason for recall
St. Jude Medical has initiated a recall of one lot of Livewire TC" Ablation Catheter. The product was mis-labelled as having an 8 mm tip instead of 4 mm tip. This could potentially result in elevated temperatures and formation of coagulum at the tip electrode. If this coagulum were to detach from the product, the worst case this could potentially result in embolization.
Action taken
Consignees were sent a "St Jude Medical Urgent Product Recall" letter addressed to "Dear Doctor" and a "Doctor's Acknowledgement Form" . The letter described the product, problem and the possibility of a health hazard. They also requested discontinue of product and gave instructions for return of product. Contact St. Jude Medical at 952-933-4700 if you have questions.
Root cause
Labeling mix-ups
Status
Terminated
Product code
LPB
Quantity affected
39
Distribution
Canada and Belgium
Date initiated
2008-06-05
Date posted
2008-09-20
Date terminated
2012-02-04
Location
Minnetonka, MN