The St. Jude Medical Livewire TC" Ablation Catheter, 8mm Tip w/Thermocouple & Thermistor REF 402196, Length 115 cm, Lot
FDA device recall Z-2313-2008 · posted 2008-09-20 · Terminated
Recall details
- Recalling firm
- St. Jude Medical Cardiovascular Division
- Reason for recall
- St. Jude Medical has initiated a recall of one lot of Livewire TC" Ablation Catheter. The product was mis-labelled as having an 8 mm tip instead of 4 mm tip. This could potentially result in elevated temperatures and formation of coagulum at the tip electrode. If this coagulum were to detach from the product, the worst case this could potentially result in embolization.
- Action taken
- Consignees were sent a "St Jude Medical Urgent Product Recall" letter addressed to "Dear Doctor" and a "Doctor's Acknowledgement Form" . The letter described the product, problem and the possibility of a health hazard. They also requested discontinue of product and gave instructions for return of product. Contact St. Jude Medical at 952-933-4700 if you have questions.
- Root cause
- Labeling mix-ups
- Status
- Terminated
- Product code
- LPB
- Quantity affected
- 39
- Distribution
- Canada and Belgium
- Date initiated
- 2008-06-05
- Date posted
- 2008-09-20
- Date terminated
- 2012-02-04
- Location
- Minnetonka, MN