The spare component Anti-HBs Cutoff Calibrator, 10 mLU/mL, Catalog number 26154. The spare component is sold separately
FDA device recall Z-1840-2015 · posted 2015-06-25 · Terminated
Recall details
- Recalling firm
- Bio-Rad Laboratories, Inc.
- Reason for recall
- The spare component Anti-HBs Cutoff Calibrator, Catalog number 26154, is recalled because the expiration date of 2016-04-30 printed on label in error.
- Action taken
- Bio-Rad Laboratories Inc, sent an Urgent Product Correction letter dated May 13, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The recalled calibrator is a spare component which is sold separately from the MONOLISA Anti-HBs EIA Kit and the kit is not affected by this recall. Customers were informed that testing was performed and the Anti-HBs 10 mLU/mL Calibrator was acceptable to use through the expiration date of the kit dated 2015-11-30. Customers with questions can call the firm at 1-800-224-6723, option #2, then #3.
- Root cause
- Incorrect or no expiration date
- Status
- Terminated
- Product code
- JJX
- Quantity affected
- 20 vials
- Distribution
- US Distribution to the states of : AZ, CA, ID, MO, and NH.
- Date initiated
- 2015-05-13
- Date posted
- 2015-06-25
- Date terminated
- 2015-11-05
- Location
- Redmond, WA