FDA device recall Z-3056-2011 · posted 2011-08-22 · Terminated

Recall details

Recalling firm
Boston Scientific Corporation
Reason for recall
Boston Scientific has received reports where, following clip deployment, the alignment of the clip and delivery system does not allow for the withdrawal of the delivery system away from the clip, resulting in tissue trauma, increased bleeding and the clip being pulled off tissue as the delivery system is withdrawn.
Action taken
Boston Scientific sent an URGENT MEDICAL DEVICE RECALL letter dated August 12, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to forward the Medical Device Recall to their customers. Customers were instructed to: Immediately discontinue use of any affected product. Complete and return the Reply Verification Tracking Form by fax to 508-683-5578. Package and return any affected product to: Boston Scientific Corporation Distribution Center Attn: QA Returns 500 Commander Shea Blvd Quincy, MA 02171 For questions regarding this recall call 508-683-5578.
Root cause
Other
Status
Terminated
Product code
FHN
Quantity affected
14, 093 boxes of both 155 cm and 235 cm size clips
Distribution
Worldwide Distribution - including USA (nationwide)
Date initiated
2011-07-19
Date posted
2011-08-22
Date terminated
2012-06-22
Location
Marlborough, MA