Atlas FDA

The product has the Catalogue/Part number 60010. Olympic Cool-Cap System is a selective head cooling system for treatmen

FDA device recall Z-0447-2013 · posted 2012-11-30 · Terminated

Recall details

Recalling firm
Natus Medical Incorporated
Reason for recall
The Olympic Cool-Cap's Control Module has experienced a frozen screen during use. When this occurs, the on-screen information remains on display, but the system is no longer providing cooling treatment to the infants. The visible clock displayed at the upper right hand corner of the screen stops advancing.
Action taken
The firm, Natus Medical Incorporated, sent its customers an "URGENT: MEDICAL DEVICE" letter dated May 16, 2012 with an attached Field Safety Notice. The letter describes the product, problem and actions taken. Customers can call Natus Technical Service at 888-496-2887 for any questions about this recall.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
MXM
Quantity affected
324 units
Distribution
Worldwide distribution: USA (nationwide) including states of: AK, AR, AZ, CA, CO, DE, FL, GA, HI, ID, IL, IN, KY, MA, MI, MN, MS, ND, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, WA, Puerto Rico, a
Date initiated
2012-05-09
Date posted
2012-11-30
Date terminated
2015-01-27
Location
Seattle, WA