The Oncology Essentials Package of the Spectral CT imaging system.
FDA device recall Z-1756-2024 · posted 2024-05-02 · Open, Classified
Recall details
- Recalling firm
- Philips North America Llc
- Reason for recall
- Potential for third party oncology marking laser to be reflected off the gantry cover surface, which can potentially lead to misdiagnosis and subsequent incorrect treatment.
- Action taken
- An URGENT Medical Device Correction Letter, dated 3/27/24, was mailed to consignees to notify them of this recall. Consignees are instructed to identify if a third-party laser was installed on their system and if the laser line is reflected, duplicated, or thickened. If a third-party laser is installed, before treatment of a patient, reposition or re-image as necessary to verify proper placement of the localization laser marks. Systems may continue to be used in accordance with the intended use and provided use recommendations. Consignees are to complete and return the provided response form by email to CTNM.QARA@philips.com. The provided recall notification is to be disseminated to all users of affected systems. Philips will contact consignees to schedule a time for a Field Service Engineer to visit and install hardware to resolve the issue. Questions are to be directed to the Customer Care Solutions Center at 1-800-722-9377 from 8:00 AM to 8:00 PM, Monday through Friday.
- Root cause
- Under Investigation by firm
- Status
- Open, Classified
- Product code
- JAK
- Quantity affected
- 12 units
- Distribution
- Worldwide distribution - US Nationwide in the states of AZ, HI, MN, VT, WV and the countries of China, Macao, New Zealand, Portugal, Switzerland.
- Date initiated
- 2024-03-29
- Date posted
- 2024-05-02
- Location
- Cambridge, MA