The Livongo Health In Touch Blood Glucose Monitoring System (BGMS), is an Over-The-Counter (OTC) system designed for the

FDA device recall Z-1587-2016 · posted 2016-04-29 · Terminated

Recall details

Recalling firm
Livongo Health
Reason for recall
The instructional insert for the In Touch Control Solution for use with the In Touch meter and In Touch strips as a quality control check did not contain the limitation of sensor accuracy: (1) with the use of acetaminophen (Tylenol) (2) with the use of Ascorbic Acid (vitamin C) supplementation (3) in the presence of uric acid
Action taken
The recall was conducted in three distinct events. Members were first sent an e-mail of the Recall Notification - Livongo Test Strip Insert (Event 1) containing information regarding the test strip insert discrepancy. Following the e-mail distribution, a hard copy of the same information was distributed (Event 2) to members via United States Postal\ Service (USPS) mail. Lastly, a second e-mail (Event 3) containing the same information was sent to members. The distribution list for the second (hardcopy distribution) and third (e-mail distribution) events was a subset of the initial member distribution.
Root cause
Error in labeling
Status
Terminated
Product code
NBW
Quantity affected
63,000 units
Distribution
US consignees: Arizona, California, Delaware, Florida, Georgia, Illinois, Kentucky, Maryland, Michigan, Minnesota, Mississippi, Missouri, Montana, New Jersey, New York, Ohio, Pennsylvania, South Carol
Date initiated
2016-02-18
Date posted
2016-04-29
Date terminated
2018-03-06
Location
Chicago, IL