Atlas FDA

The GemStar Power Supply is an accessory for the GemStar Infusion Pump. It is a two piece assembled switching power supp

FDA device recall Z-0095-2015 · posted 2014-10-24 · Terminated

Recall details

Recalling firm
Hospira Inc.
Reason for recall
The GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one instance, smoke was observed and the GemStar pump was operating on battery power while connected to the 3 volt DC power supply.
Action taken
An URGENT MEDICAL DEVICE CORRECTION letter dated February 3, 2014 was sent to all direct accounts. The letter included specific instructions for how to handle and use the power supplies so as to minimize the chance of any device failures. There were additional instructions for customers to: 1) Inform potential users in their organization of the product of this notification; 2) Complete the attached reply form and return it to the fax number or e-mail address on the form (even if you do not have any affected product in inventory); 3) If the products were further distributed, their customers should be notified of the recall and asked to contact Stericycle at 888-641-9735 to receive a reply form; and, 4) if a potential user believes that their power supply is damaged for any reason, it should be removed from service and the user should contact Hospira Global Complaints at 800-441-4100 to report the concern. The letter further notes that there is no need to return fully functional power supplies (unless they malfunction).
Root cause
Process control
Status
Terminated
Product code
FRN
Quantity affected
978 power supplies
Distribution
Worldwide Distribution -- USA, including the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, P
Date initiated
2014-02-03
Date posted
2014-10-24
Date terminated
2017-03-07
Location
Lake Forest, IL

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
HOSPIRA GEMSTAR INFUSION PUMP SYSTEM, 7 THERAPY, MODEL 13087; 6 THERAPY, MODEL 13086; PAIN MANAGEMENT, MODEL 13088 (K060806)Hospira, Inc.2006-04-21