The FastPack IP Sex Hormone Binding Globulin Immunoassay, Part #25000081, is for the quantitative measurement of SHBG in

FDA device recall Z-0805-2025 · posted 2025-01-03 · Open, Classified

Recall details

Recalling firm
Qualigen Inc
Reason for recall
Three lots of the FastPacks kitted into the FastPack SHBG Complete Kits were labeled with the incorrect expiration date. The following is a listing of affected lots: Lot Number: 2408018-1, 2409015-1, and 2409015-2. The likelihood of patient harm is expected to be low as routine Quality Control should detect kits that are impacted by signal deterioration due to age.
Action taken
The recalling firm issued letters dated 11/19/2024 via email to their customers explaining the reason for recall and providing the correct expiration dates for each of the affected lot numbers. The risk to health was provided and instructions given to be completed. The instructions were to cease use of the affected product, dispose of all remaining kits from the affected lot numbers, document the number of kits disposed, complete the acknowledgment form by FAX or email within 5 days.
Root cause
Labeling Change Control
Status
Open, Classified
Product code
CDZ
Quantity affected
426
Distribution
Distribution was made to AR, CA, CO, FL, IL, IN, MA, MO, NC, NY, OH, OK, TN, TX, and VA. There was no military/government/foreign distribution.
Date initiated
2024-12-03
Date posted
2025-01-03
Location
Carlsbad, CA