The Evolution steam sterilizers, manufactured by Steris Corporation, utilize a loading system to assist the operator in

FDA device recall Z-1522-2012 · posted 2012-05-11 · Terminated

Recall details

Recalling firm
Steris Corporation
Reason for recall
Through Customer feedback and field service experience, STERIS learned that operators may experience difficulty when inserting or removing equipment from the Evolution sterilizer using the transfer carriage due to the following issues with the transfer carriage. 1) Deformation of the latch pin or cracked ball bearings preventing the roller from turning. 2) The rollers on the ends of the pivot a
Action taken
Steris Corporation sent an Urgent Voluntary Field Correction Notice dated March 23, 2012 via FedEx with tracking numbers for delivery confirmation to each affected customer. The letter identified the affected product, problem and actions to be taken. The letter stated a STERIS Field Service Representatives will contact each affected customer to schedule a service visit to upgrade the transfer carriages. For technical questions or questions regarding STERIS visit to your facility, please contact STERIS Field Service Dispatch at 1-800-333-8828.
Root cause
Device Design
Status
Terminated
Product code
FLE
Quantity affected
523
Distribution
Worldwide Distribution - USA Nationwide including AL, AZ, CA, CO, CN, DE, FL, GA, IL, IN, KA, KY, LA, MD, MA, MI, ME, MO, MT, NE, NH, NJ, NM, NY, NC, ND, OH, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, &
Date initiated
2012-03-23
Date posted
2012-05-11
Date terminated
2013-06-11
Location
Mentor, OH