The Evolution steam sterilizers, manufactured by Steris Corporation, utilize a loading system to assist the operator in
FDA device recall Z-1522-2012 · posted 2012-05-11 · Terminated
Recall details
- Recalling firm
- Steris Corporation
- Reason for recall
- Through Customer feedback and field service experience, STERIS learned that operators may experience difficulty when inserting or removing equipment from the Evolution sterilizer using the transfer carriage due to the following issues with the transfer carriage. 1) Deformation of the latch pin or cracked ball bearings preventing the roller from turning. 2) The rollers on the ends of the pivot a
- Action taken
- Steris Corporation sent an Urgent Voluntary Field Correction Notice dated March 23, 2012 via FedEx with tracking numbers for delivery confirmation to each affected customer. The letter identified the affected product, problem and actions to be taken. The letter stated a STERIS Field Service Representatives will contact each affected customer to schedule a service visit to upgrade the transfer carriages. For technical questions or questions regarding STERIS visit to your facility, please contact STERIS Field Service Dispatch at 1-800-333-8828.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- FLE
- Quantity affected
- 523
- Distribution
- Worldwide Distribution - USA Nationwide including AL, AZ, CA, CO, CN, DE, FL, GA, IL, IN, KA, KY, LA, MD, MA, MI, ME, MO, MT, NE, NH, NJ, NM, NY, NC, ND, OH, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, &
- Date initiated
- 2012-03-23
- Date posted
- 2012-05-11
- Date terminated
- 2013-06-11
- Location
- Mentor, OH
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| AMSCO EVOLUTION MEDIUM STEAM STERILIZER, MODEL HC-900 AND HC-1200 (K091136) | STERIS Corporation | 2009-06-29 |
| AMSCO EVOLUTION MEDIUM STEAM STERILIZER (K082435) | STERIS Corporation | 2008-12-30 |