The Dimension FT4L Flex reagent cartridge( Free Thyroxine) manufactured by Siemens Healthcare Diagnostics, Inc., Newark,

FDA device recall Z-0458-2011 · posted 2010-11-24 · Terminated

Recall details

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Reason for recall
Some LOCI cartridges have sub-optimal seal of the lidstock that can cause rupture of the package at high elevations, causing the cartridge to leak. Use of leaking cartridges may result in inaccurate test results.
Action taken
Siemens issued an "Urgent Field Safety Notice" dated April 2010, informing users of the problem and instructing them to inspect cartridges for leaking fluid. Cartridges with leaking fluid should be reported to the Siemens Technical Solutions Center. The Technical Solutions Center can be contacted at 800 441 9250.
Root cause
Other
Status
Terminated
Product code
CEC
Quantity affected
1520
Distribution
Worldwide Distribution: USA, Europe, Australia, China, Japan, Korea, and Saudi Arabia.
Date initiated
2010-04-12
Date posted
2010-11-24
Date terminated
2013-03-19
Location
Newark, DE