Atlas FDA

The device is the NeuViz Dual series Computed Tomography (CT) Scanner System. The device is identified as NeuViz Dual, P

FDA device recall Z-0846-2011 · posted 2011-01-04 · Terminated

Recall details

Recalling firm
Philips And Neusoft Medical Systems Co., Ltd.
Reason for recall
Potential for the CT patient table top to become detached from the carrier due to missing washers.
Action taken
Neusoft Medical Systems Co., China sent an URGENT DEVICE CORRECTION letter dated Feb. 10, 2010 to their distributor in NC. On 02/12/10, The distributor sent the same letter to their consignees. The firm informed this distributor that there was report of the four screws that secure the patient table top to the carrier that did not have the proper washers in place which caused the table top to detach. If the issue would reoccur, there is a potential for the table top to fall and cause patient injury. Customers were instructed to check the patient table top fixing. If it is found that the the assembly is improper customers were instructed to top using the product and contact local Service personnel for repair. Any customer can contact Service Support Department, Neusoft Medical Systems Co, LTD at nms-service@neusoft.com or Service Support Department at helpdesk@neusoft.com for questions about this recall.
Root cause
Other
Status
Terminated
Product code
JAK
Quantity affected
26 devices
Distribution
Nationwide Distribution including NC
Date initiated
2010-02-10
Date posted
2011-01-04
Date terminated
2011-04-15
Location
Shenyang, Liaoning, China

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
MODIFICATION TO NEUVIZ DUAL MULTI-SLICE CT SCANNER SYSTEM (K071308)Neusoft Medical Systems Co., Ltd.2007-08-16
NEUVIZ DUAL MULTI-SLICE CT SCANNER SYSTEM (K062451)Neusoft Medical Systems Co., Ltd.2006-09-28