The CEP 8 SpectrumGreen (SG) ASR Probe Kit, 20ul (list 06J37-018) is an Analyte Specific Reagent (ASR) containing a sing
FDA device recall Z-1491-2014 · posted 2014-04-23 · Terminated
Recall details
- Recalling firm
- Abbott Molecular
- Reason for recall
- 20 kits of CEP 8 SG ASR Probe (list 06J37-018, lot 440023), contained Vysis CEP 4 (Alpha Satellite) SG Probe, 20u, (part 32-112004 lot 438814), instead of the expected CEP 8 probe.
- Action taken
- Customers were sent Field Correction Notices dated 3/28/13 instructing them to inspect their CEP 8 SpectrumGreen Probe kits for the presence of CEP4 probe vials. If any CEP4 vials are discovered, customers should contact their local Abbott Molecular Technical Support Represtative for coordination of reagent replacement with the correct CEP8 vial. If customers have any specific questions they can contact Abbott Molecular at 1-800-553-7042, Option 2.
- Root cause
- Packaging change control
- Status
- Terminated
- Product code
- MVU
- Quantity affected
- 20 kits
- Distribution
- Nationwide (CA, CO, CT, FL, HI, IL, IN, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, TN, TX, VA, WA, WI); Canada
- Date initiated
- 2013-03-28
- Date posted
- 2014-04-23
- Date terminated
- 2016-03-08
- Location
- Des Plaines, IL