The Brilliance Workspace Portal is a software component of the Computed Tomography X-Ray System. The Brilliance Workspac
FDA device recall Z-1219-2012 · posted 2012-03-16 · Terminated
Recall details
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Reason for recall
- Incorrect Standardized Uptake Value ("SUV") values are displayed in two instances: 1. After loading PET data (with a Slope value different than 1) into the Quick Review (QR) application from a remote device; and 2. During the loading the above mentioned type of PET data into the CT viewer application from a remote device.
- Action taken
- The Philips Healthcare, sent an "URGENT-Medical Device Correction" Field Safety Notice dated February 08, 2012 to their customers. The letter describes the product, problem and actions to be taken. The customers were instructed to review the information with all members of their staff and to retain a copy with the equipment Instruction for Use. The letter notifies the customers that a Field Service Engineer will be contacting them to schedule a visit to implement the corrective action by installing the software update version 2.6.1. If you need further information or support concerning this issue, please contact your local Philips representative or local Philips Healthcare office. For North America and Canada contact the Customer Care Solutions Center (1-800-722-9377, option 5: Enter Site ID or follow the prompts).
- Root cause
- Software design
- Status
- Terminated
- Product code
- KPS
- Quantity affected
- 822
- Distribution
- Worldwide distributions: USA (nationwide) including the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, VA, NY, OH, OK, OR, P
- Date initiated
- 2012-02-23
- Date posted
- 2012-03-16
- Date terminated
- 2013-02-22
- Location
- Cleveland, OH