The Bigger Better-Bladder, Catalog Number BBB38 (with 3/8 inch ID tubing), Sterile, Single use only --- Manufactured for
FDA device recall Z-1323-2010 · posted 2010-04-13 · Terminated
Recall details
- Recalling firm
- Circulatory Technology Inc
- Reason for recall
- During an FDA inspection, it was discovered that the firm had conducted a silent recall of its Bigger Better Bladder device after receiving a complaint from a hospital that the device collapsed during a procedure thereby obstructing flow.
- Action taken
- The firm, Circulatory Technology, Inc., sent an email to Bigger Better-Bladder end user accounts on 6/03/09 and again on 6/13/09. The firm followed up with phone calls to the user customers. The emails and calls discussed the product, problem and the actions to be taken by the customers. The customers are ask to examine their stock of Bigger Better-Bladder for the affected units and then visually examine each unit to assure that adhesive has been applied and when the tube is bent away from cap, no separation is visible along the joint between the two according to the picture provided. The customers should return for full credit or a replacement of any product that has a separation of the seal. Please feel free to contact Mr. Yehuda Tamari at 516-624-2424 with any questions.
- Root cause
- Process control
- Status
- Terminated
- Product code
- DTN
- Quantity affected
- 144 units
- Distribution
- Arizona, Arkansas, California, Florida, Indiana, Iowa, Maine, Michigan, Nebraska, New York, North Carolina, Pennsylvania, Ohio, Texas, Utah, Washington DC, and Germany.
- Date initiated
- 2009-06-03
- Date posted
- 2010-04-13
- Date terminated
- 2010-05-14
- Location
- Oyster Bay, NY
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| BETTER-BLADDER (BB) (K981284) | Circulatory Technology, Inc. | 1998-08-12 |