Atlas FDA

The Alaris PC Unit, Model Numbers 8000 and 8015 (formerly Medley PC Unit) software version 8 and above when used with th

FDA device recall Z-1791-2009 · posted 2009-08-03 · Terminated

Recall details

Recalling firm
Cardinal Health 303 dba Cardinal Health
Reason for recall
Affected Unit: Alaris PC Unit (Models 8000 and 8015) with software a versions 8 and above when used with the Alaris PCA module (Model 8120) There is a potential risk of over infusion related to a warning message on the user interface of the Alaris PC unit when used in conjunction with the Alaris PCA module. When the Alaris PC unit detects a programmed infusion for the Alaris PCA module that exceed
Action taken
The recall was initiated on June 12, 2009 with the firm forwarding via UPS an Urgent: Medical Device Recall Notification with attached Customer Response Card to the Chief Administrative Officer, Director of Pharmacy, Director of Biomedical Engineering, Director of Nursing & Director of Risk Management of all of the affected consignees. Customers were also asked to please promptly complete and return the enclosed Customer Response Card to expedite the correction process. The Recall Notification informed the consignees of: (A) PROBLEM #1: Occlusion Warning Message for Alaris Pump Module; (B) PROBLEM #2 Syringe volume Warning Message for Alaris PCA module; (C) PROBLEM #3 Electrostatic Discharge (ESD) protection circuitry; (D) PROBLEM #4 Fluid Ingress and (E) PROBLEM #5 Alaris System Inter-Unit Interface (IUI) Connectors Labeling. The recall notice included "Required Action for Users & Follow-up actions by Cardinal Health for all five problems. Customers were also asked to report any adverse reactions experienced with the use of this product, and/or quality problems to the FDA's MedWatch Program. In addition, customer were provided the following contact information for the recall: (1) Cardinal Health Recall Center for general questions: By phone at 1-888-562-6018 7am to 5pm (Pacific) ; (2) Customer Advocacy for adverse reports: By phone at 1-800-854-7120, Option 1, Option 1, Option 3 or by email at customerfeedback@cardinalhealth.com 24 hours a day, Sunday-Saturday; (3) Technical Support for Technical Questions regarding the Alaris System: By phone at 1-888-562-6018, 6am-5pm (Pacific). The firm followed-up with the release of a nationwide press statement regarding this issue on July 29, 2009. The press release indicated serial numbers of affected devices, as well as CareFusions short term instructions to customers, and the firms strategy to fix the affected devices can be found at: www.cardinalhealth.com/alaris/medical-device-recall/ (This informatio
Root cause
Device Design
Status
Terminated
Product code
FRN
Quantity affected
186,000 units
Distribution
Worldwide Distribution -- US (including Puerto Rico and Guam), Germany, England, Canada, Netherlands, Malaysia, Singapore, Australia, and New Zealand.
Date initiated
2009-06-12
Date posted
2009-08-03
Date terminated
2012-03-19
Location
San Diego, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ALARIS SYSTEM PC UNIT, MODEL 8001 (K051641)Cardinal Health, Alaris Products2005-10-20
GEMINI PC-1,PC-2, PC-2TX,PC-4, SIGNATURE EDITION INFUSION PUMP (SE), MEDSYSTEM III INFUSION (MSIII), MEDSYSTEM PATIENT C (K012383)Alaris Medical Systems, Inc.2001-10-02