The Abbott m2000sp is an automated system for performing sample preparation for nucleic acid testing.
FDA device recall Z-0948-2015 · posted 2015-01-09 · Terminated
Recall details
- Recalling firm
- Abbott Molecular
- Reason for recall
- The x-drive assembly may fail prematurely. This could cause the manipulator or liquid handling arm to contact an object, shake and drip, or improperly pipet.
- Action taken
- Abbott Molecular sent a different Field Correction Recall / Urgent Field Safety Notice letters dated October 24, 2014 to each of three sets of customers. Customers whose instruments' X-Drive Assembly belts were perviously inspected by a Field Service Representative and found to not require replacement were advised to contact their local Abbott Molecular Representative or Abbott Molecular Technical Support at 800-553-7042 x 2 should they hear any unusual noises associated with the LiHa or RoMa movement or misalignment of the LiHa or RoMa. Customers whose instruments' X-Drive Assembly belts were previously inspected and replaced by a Field Service Representative were notifieied that this maintenance activity took place. Customers with questions were instructed to contact Abbott Molelcular. For questions regarding this recall call 224-361-7000. Customers whose instruments' X-Drive Assembly belts have not been inspected were notified that a Field Service Representative is scheduled to visit their facility to inspect and if necessary, replace the X-Drive Assembly belt(s). Customers were advised to contact their local Abbott Molecular Representative should they hear any unusual noises associated with the LiHa or RoMa movement or misalignment of the LiHa or RoMa.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- JJH
- Quantity affected
- 176 devices
- Distribution
- Worldwide Distribution - USA including CA, FL, GA, IL, MI, NC, NJ, NY, OH, OK, OR, and TX and Internationally to Australia, Austria, Belgium, Bosnia And Herzegovina, Brazil, Canada, China, Colombia, C
- Date initiated
- 2013-11-23
- Date posted
- 2015-01-09
- Date terminated
- 2017-12-04
- Location
- Des Plaines, IL