Test Request Distributor (TRD 1.3 and TRD 1.4), a software module used on the Dako Omnis System, an automated slide stai
FDA device recall Z-1548-2015 · posted 2015-04-29 · Terminated
Recall details
- Recalling firm
- Dako North America Inc.
- Reason for recall
- If a user requests slides from the LIS or TPID, then updates a request by changing the test, the TRD software will reject the update. The LIS and printed slide label will display the updated test with the requested change, but the Omnis system will execute the initial test, not reflecting the change. The system will not warn the user that the test request was rejected.
- Action taken
- Dako sent a Voluntary Product Recall Notification letter starting the week of April 1, 2015 along with a acknowledgement attachment. The letter identified the affected product, problem and actions to be taken. The firm requests that customers inform those in their organization of the notification including organizations where the potentially affected instruments have been transferred. The firm requests that customers complete, sign, and return the acknowledgement form within one week after receiving the notification. For questions contact your Dako sales representative. On 04/29/15 the firm sent out an updated customer notification letter in order to provide additional clarification and instructions to customers in regards to the recall situation.
- Root cause
- Other
- Status
- Terminated
- Product code
- KPA
- Quantity affected
- 33 (7 in U.S. and 26 Internationally)
- Distribution
- Worldwide Distribution - U.S Nationwide in the states of IN and PA and the countries of: Spain, Belgium, France, Netherlands, Denmark, Finland, Sweden, Germany, UK, and Korea.
- Date initiated
- 2015-04-01
- Date posted
- 2015-04-29
- Date terminated
- 2015-08-05
- Location
- Carpinteria, CA