Tessier Osseous Microtome TOM(R) Bone Mill, 01-15402, Stryker Portage, MI. The TOM bone mill is designed to grind any ty

FDA device recall Z-1202-2010 · posted 2010-03-23 · Terminated

Recall details

Recalling firm
Stryker Craniomaxillofacial Division
Reason for recall
The firm has received 328 customer complaints for this device. Most of the complaints were involving non-functioning of the ratcheting handle or complete device, or the presence of metal shavings or discoloration.
Action taken
An "URGENT MEDICAL DEVICE CORRECTION" letter dated February 18, 2010, was sent to the customers. The letter describes the product, problem, scope of affected products, potential patient risks and actions to be taken by the customers. The customers should review the letter with all users of the TOM(r) Bone Mill and follow the instructions on the proper inspection and maintenance of the device enclosed with the letter. The customers should fax the accountability form within 5 days to help Stryker ensure their receipt of the information. If the accountability form is not received, addition communication will be sent. If you have any urgent questions or concerns, please contact us at (269) 323-4258. Please ref: PFA #2009-128. Our normal business hours are Monday-Friday 8am-5pm (GMT+1:00) On 7/8/2010 the firm sent an Urgent Medical Device Recall letter to its customers. The letter stated thatt upon consultation with regulatory agencies outside of the United States, the firm had intiated a devicd removal. The letter instructed the customers to return any devices that they had to the recalling firm.
Root cause
Equipment maintenance
Status
Terminated
Product code
LXH
Quantity affected
596 units
Distribution
Worldwide distribution: USA and Belarus, Bulgaria, Denmark, Finland, Hungary, Jordan, Norway, Russia, Saudi Arabia and the UAE.
Date initiated
2010-02-18
Date posted
2010-03-23
Date terminated
2010-11-24
Location
Portage, MI