Atlas FDA

Terumo Sarns Level Sensor II (yellow) (Part # 195215) alert level sensor transducer for use with Sarns Advanced Perfusio

FDA device recall Z-0295-2009 · posted 2008-11-07 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corporation
Reason for recall
The level sensor may not properly couple to the reservoir, resulting in a sensor "not attached" message, an alert or alarm condition, or a failure to detect a low level condition.
Action taken
U.S. consignees were notified by (Urgent Medical Device Removal) letter dated 5/23/08 via FedEx and international affiliates were notified via e-mail on 5/27/08. Consignees were advised of the problem, and to continue using the affected sensors until replacements are available in late 2008. For questions or additional information, contact the Terumo CVS Customer Service at 1-800-521-2818.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
DTQ
Quantity affected
574
Distribution
Worldwide including USA and countries of Australia, Belgium, Bolivia, Canada, Chile, Ecuador, Egypt, Honduras, Hong Kong, India, Indonesia, Japan, Jordan, Latin America, Malaysia, Panama, Philippines,
Date initiated
2008-05-23
Date posted
2008-11-07
Date terminated
2009-08-04
Location
Ann Arbor, MI