Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly co

FDA device recall Z-1116-2015 · posted 2015-02-13 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corporation
Reason for recall
Malfunctioning mixing valve and loss of cooling and/or heating capability. System goes to standby mode and shows error messages EOd (invalid mix valve measured range of motion) or EOE (valve unable to move or complete a motion). Mixing valves need to be replaced at regular intervals.
Action taken
On 2/6/2015, URGENT MEDICAL DEVICE RECALL notifications were mailed to the affected hospitals. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call customer service at 1-800-521-2818.
Root cause
Device Design
Status
Terminated
Product code
DWC
Quantity affected
103 units
Distribution
US (nationwide) including the states of AR, CA, CO, CT, FL, IA, IL, KS, KY, MD, MI, MO, MS, NY, OH, PA, TX, UT, WA, and WI and the country of Japan.
Date initiated
2015-02-06
Date posted
2015-02-13
Date terminated
2016-10-17
Location
Ann Arbor, MI