Atlas FDA

Terumo HX2 Temperature Management System, Catalog No. 809810 The Terumo HX2 Temperature Management System is indicated f

FDA device recall Z-1416-2018 · posted 2018-04-18 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corporation
Reason for recall
Update to cooler-heater cleaning instructions.
Action taken
The firm,Terumo Cardiovascular Systems Corporation, mailed customers an "Urgent Medical Device Correction" letter dated 2/16/18. The letter described the product, problem and actions to be taken. The Customers were instructed to do the following: 1) discard the current Sarns TCM, TCM II Cooling and Heating, or tHX2 Temperature Management Systems cleaning guides; 2) Review the notice and updated cleaning guides, and ensure all users receive notice of the updated cleaning guides and; 3) Return the Customer Response Form acknowledging receipt of the notice via Terumo Recall Email: tcvs.recall@terumomedical.com or Terumo Recall Fax: 734-741-6149. The letter also includes a list of the updates to the cleaning guides. Customers with questions may contact Terumo CVS Customer Service (800) 521-2818, Monday-Friday, 8am-6pm ET or Fax (734)741-6149.
Root cause
Labeling design
Status
Terminated
Product code
DWC
Quantity affected
2393 this total is for all products
Distribution
Worldwide Distribution: US (Nationwide), Australia, Belgium, Canada, Chile, China. Colombia, Dominican Republic, England, Finland, France, Germany, Greece, HK, India, Indonesia, Iran. Israel, Italy, J
Date initiated
2018-03-16
Date posted
2018-04-18
Date terminated
2020-06-10
Location
Ann Arbor, MI

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
TERUMO HX2 TEMPERATURE MANAGEMENT SYSTEM (K071521)Terumo Cardiovascular Systems Corp.2007-11-21