Terumo Cardiovascular Systems Corporations, CDI 100 Hematocrit/ Oxygen Saturation Monitoring System. In-line blood-gas s

FDA device recall Z-1944-2011 · posted 2011-04-11 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corporation
Reason for recall
Inaccurate result. Possibility for displayed values for potassium, CO2, hematocrit, O2 saturation, pH, hemoglobin and other measured blood parameters to be inaccurate.
Action taken
On 11/1/2010, Terumo sent "URGENT MEDICAL DEVICE CORRECTION" notices to Chief Perfusionist, Department of Cardiovascular Surgery or Director of Operating Room Services. The letter informed the consignees about the incidents of the CDI 100 and 101 displaying inaccurate values. The letter states that when a displayed value is significantly different from expectations based on the clinical situation, the user should verify the accuracy by independent means before treating. The letter included "Caution" stickers that were to be applied to the CDI system operators manual.
Root cause
Device Design
Status
Terminated
Product code
DTY
Quantity affected
1032 units
Distribution
US, India, Colombia, Taiwan, Venezuela, Vietnam, Puerto Rico, Malaysia, Sri Lanka, Brazil, Indonesia, Pakistan, Chile, Germany, Hong Kong, Japan, AE, Singapore, Austria, Belgium and Thailand.
Date initiated
2010-11-01
Date posted
2011-04-11
Date terminated
2012-04-17
Location
Ann Arbor, MI