Terumo Cardiovascular Systems, CDI blood parameter monitoring system 500 with venous blood parameter module, catalog 500
FDA device recall Z-2682-2011 · posted 2011-06-28 · Terminated
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Reason for recall
- Inaccurate readings may result after Methylene Blue, an intravascular dye, is administered to the patient. The presence of Methylene Blue may interfere with the readings from the shunt sensor for the CDI 500 system in a way that would cause inaccurate pH values. Because the pH value is used to calculate other values, these parameters could also be affected by the presence of Methylene Blue: K
- Action taken
- Terumo sent URGENT MEDICAL DEVICE CORRECTION letters to Chief of Perfusion, Department of Cardiothoracic Surgery or Director of Operating Room Services, dated 4/21/2011. The letter explained the reason for the correction and the potential hazard. The letter stated that Terumo was updating the Operators Manual for all CDI systems and would provide the manuals to all users when available. Consignees with questions should call 1-800-521-2818. Monday- Friday 8 am- 6 pm.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- DRY
- Quantity affected
- 3767 all CDI 500 models
- Distribution
- Worldwide distribution, including USA, Australia, Barbados, BELGIUM, BRAZIL, CANADA, Chile, Colombia, Costa Rica, Dominican Republic, ECUADOR, El Salvador, France, Germany, Guatemala, Holland, Hondura
- Date initiated
- 2011-04-21
- Date posted
- 2011-06-28
- Date terminated
- 2012-08-06
- Location
- Ann Arbor, MI