Terumo Backplate Replacement CXFVR 1500 Holder The Flexible venous Reservoir is mounted on the Backplate/Holder and secu
FDA device recall Z-0071-2013 · posted 2012-10-16 · Terminated
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Reason for recall
- Backplate reservoir holders were manufactured with an incorrect configuration of connector supports. The devices were manufactured and configured with two 3/8 inch and one inch connector supports instead of one 3/8 inch and two inch connector supports (See location of supports as # 7 and # 8 on backplate attachment). The backplates were manufactured as described in manufacturing instructions a
- Action taken
- The firm, Terumo, interviewed the Sr. Product Manager from the Marketing Department of the company on August 29, 2012. The interview included the description of the product, problem and action taken. Replacement connectors were provided to users and all units were corrected. If you have any questions, call 734-663-4145 or Manager, Regulatory Affairs at 734-741-5576.
- Root cause
- Labeling design
- Status
- Terminated
- Product code
- KRI
- Quantity affected
- 33 units
- Distribution
- Worldwide distribution: USA (nationwide) including states of: CA, MI, NC, NJ, PA, and WI;and country of: EUROPE.
- Date initiated
- 2009-10-02
- Date posted
- 2012-10-16
- Date terminated
- 2013-01-29
- Location
- Ann Arbor, MI