Terumo Backplate Replacement CXFVR 1500 Holder The Flexible venous Reservoir is mounted on the Backplate/Holder and secu

FDA device recall Z-0071-2013 · posted 2012-10-16 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corporation
Reason for recall
Backplate reservoir holders were manufactured with an incorrect configuration of connector supports. The devices were manufactured and configured with two 3/8 inch and one inch connector supports instead of one 3/8 inch and two inch connector supports (See location of supports as # 7 and # 8 on backplate attachment). The backplates were manufactured as described in manufacturing instructions a
Action taken
The firm, Terumo, interviewed the Sr. Product Manager from the Marketing Department of the company on August 29, 2012. The interview included the description of the product, problem and action taken. Replacement connectors were provided to users and all units were corrected. If you have any questions, call 734-663-4145 or Manager, Regulatory Affairs at 734-741-5576.
Root cause
Labeling design
Status
Terminated
Product code
KRI
Quantity affected
33 units
Distribution
Worldwide distribution: USA (nationwide) including states of: CA, MI, NC, NJ, PA, and WI;and country of: EUROPE.
Date initiated
2009-10-02
Date posted
2012-10-16
Date terminated
2013-01-29
Location
Ann Arbor, MI