Terumo Advanced Perfusion System 1 (with 4 inch diameter roller pump); Catalog number 801040.

FDA device recall Z-0666-05 · posted 2005-03-31 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corp
Reason for recall
The pumps will stop, show an overspeed error message, and will not restart using routine procedures if the user reduces the pump speed to zero RPM while the pump is operating in pulse mode.
Action taken
Recall letters dated 3/17/05 were sent to customers advising them of the problem, the temporary fix, and that software will be upgraded.
Root cause
Other
Status
Terminated
Product code
DTQ
Distribution
Nationwide, Australia, Belgium, Canada, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Malasia, Mexico, Pakistan, South Africa, South Korea, Taiwan, Thailand and United Arab Emirates.
Date initiated
2005-03-17
Date posted
2005-03-31
Date terminated
2005-09-16
Location
Ann Arbor, MI