Terumo Advanced Perfusion System, 1 Roller Pump, 6 inch diameter; Catalog No. 801041, Terumo Cardiovascular Systems, Ann
FDA device recall Z-0322-2008 · posted 2007-12-01 · Terminated
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corp
- Reason for recall
- Failure to Operate: The pump may fail to operate, and be difficult to hand crank, due to an internal electrical short condition.
- Action taken
- Consignees were notified via an Urgent Medical Device Correction letter dated 9/10/07, which provided them guidance in the event the problem occurs and advised them they would be contacted to schedule replacement of the circuit boards.
- Root cause
- Process control
- Status
- Terminated
- Product code
- DTQ
- Quantity affected
- 76
- Distribution
- Worldwide Distribution-USA including states of Alabama, Arizona, Florida, Illinois, Indiana, Mississippi, Nebraska, North Carolina, Pennsylvania, and Virginia, and countries of Belgium, Hong Kong and
- Date initiated
- 2007-09-10
- Date posted
- 2007-12-01
- Date terminated
- 2008-02-23
- Location
- Ann Arbor, MI