Terumo Advanced Perfusion System 1 Roller Pump; 6 inch diameter; Catalog number 801040.
FDA device recall Z-0103-04 · posted 2003-11-07 · Terminated
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corp
- Reason for recall
- The roller pump tube clamp mechanism may fail to release , making it difficult to remove or insert the tubing, and thereby delaying perfusion.
- Action taken
- Letters entitled Urgent Safety Alerts were sent to each customer on 10/16/03. Customers were made away of the problem and advised to develop a protocol for action, in case the problem occurred in their facility.
- Root cause
- Other
- Status
- Terminated
- Product code
- DTQ
- Quantity affected
- 263
- Distribution
- United States, Belgium, Canada, Dubai, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Singapore, South Korea.
- Date initiated
- 2003-10-16
- Date posted
- 2003-11-07
- Date terminated
- 2005-07-07
- Location
- Ann Arbor, MI