Terumo Advanced Perfusion System 1 Flow Module; Catalog number 802018.

FDA device recall Z-0333-2007 · posted 2006-12-30 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corp
Reason for recall
Device module(s) may fail to properly initialize during start up, thus resulting in that single module being inoperable and a red 'X' being displayed on the monitor screen.
Action taken
U.S. consignees were sent an addendum to the operators manual on October 12, 2004 instructing them to cycle power to the unit to clear the error. International distributors were requested to provide this information to their customers on 12/15/04, but Terumo has not yet received confirmation that this was completed.
Root cause
Other
Status
Terminated
Product code
DTQ
Quantity affected
4740 for all affected products
Distribution
Worldwide Distribution-USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South
Date initiated
2004-10-12
Date posted
2006-12-30
Date terminated
2007-02-01
Location
Ann Arbor, MI