Terumo Advanced Perfusion System 1 Central Control Monitor. Terumo Cardiovascular Systems, Ann Arbor, MI; Catalog number
FDA device recall Z-1224-2009 · posted 2009-05-29 · Terminated
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Reason for recall
- The monitor may exhibit start-up failure, video/display issues, calibration issues, and partial or complete loss of functionality during use.
- Action taken
- Consignees were notified via an Urgent Medical Device Correction: Safety Advisory letter dated 1/12/09 advised of the reason for the correction, potential problems, and provided clinical information should the problem(s) arise. Customers are to review the notice, assure that all users are aware of the notice, and confirm receipt of the communication by completing the attached Customer Response Form and faxing it to the number indicated on the form. Questions or concerns should be directed to Terumo CVS Customer Service at 1-800-521-2818.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- DTQ
- Quantity affected
- 951
- Distribution
- Worldwide Distribution -- USA, Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Honduras, Hong Kong, India, Indonesia, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Ru
- Date initiated
- 2009-01-12
- Date posted
- 2009-05-29
- Date terminated
- 2009-07-14
- Location
- Ann Arbor, MI