Terumo Advanced Perfusion System 1 Central Control Monitor. Terumo Cardiovascular Systems, Ann Arbor, MI; Catalog number

FDA device recall Z-1224-2009 · posted 2009-05-29 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corporation
Reason for recall
The monitor may exhibit start-up failure, video/display issues, calibration issues, and partial or complete loss of functionality during use.
Action taken
Consignees were notified via an Urgent Medical Device Correction: Safety Advisory letter dated 1/12/09 advised of the reason for the correction, potential problems, and provided clinical information should the problem(s) arise. Customers are to review the notice, assure that all users are aware of the notice, and confirm receipt of the communication by completing the attached Customer Response Form and faxing it to the number indicated on the form. Questions or concerns should be directed to Terumo CVS Customer Service at 1-800-521-2818.
Root cause
Device Design
Status
Terminated
Product code
DTQ
Quantity affected
951
Distribution
Worldwide Distribution -- USA, Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Honduras, Hong Kong, India, Indonesia, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Ru
Date initiated
2009-01-12
Date posted
2009-05-29
Date terminated
2009-07-14
Location
Ann Arbor, MI