Terumo Advanced Perfusion System 1- 6" Roller Pump for Terumo System 1 (gray) Product Usage: The Terumo Advanced Perfusi

FDA device recall Z-0300-2012 · posted 2011-11-29 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corporation
Reason for recall
Terumo CVS discovered that our In house service center has incorrectly installed a felt washer into the main bearing of nine roller pumps that had been returned for repair or preventative maintenance unrelated to the felt washer. The felt washer was added as a validated running change to production to all roller pumps manufactured since February 2009. The felt washer should not have been installed
Action taken
Terumo sent an URGENT MEDICAL DEVICE REMOVAL notification letter dated November 10, 2011 to all affected customers. The letter described the affected products, problem, potential hazards, and important clinical information. Replacement roller pumps will be configured at the customer site by a Terumo Field representative with the affected roller pumps being returned via normal returned goods process. Customers were asked to complete the attached response form and fax to the number indicated on the form. For questions or concerns contact Terumo CVS Customer Service at 1-800-521-2818.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
DTQ
Quantity affected
1 unit
Distribution
Worldwide Distribution - USA (nationwide) including the states of: CA, KY, MO, NE, PA and TX and the countries of Canada and Europe.
Date initiated
2011-11-10
Date posted
2011-11-29
Date terminated
2012-09-06
Location
Ann Arbor, MI