Terumo Advanced Perfusion System 1 (4 inch diameter Roller Pump, Gray; Model 816570). Extracorporeal circulation of bloo
FDA device recall Z-0356-2009 · posted 2008-12-11 · Terminated
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Reason for recall
- Overspeed and underspeed errors, erratic pump behavior, jerky operation at low RPMs, pump instability, pump slowdowns and pump stops due to overspeed max events may occur due to a grease leak onto the motor speed encoder disk.
- Action taken
- Consignees were notified of the problem via recall letter dated 7/18/08. The firm began conducting field upgrades on the roller pumps in 07/08. For questions or concerns, contact Terumo CVS Customer Service at 1-800-521-2818.
- Root cause
- Component design/selection
- Status
- Terminated
- Product code
- DTQ
- Quantity affected
- 269
- Distribution
- Worldwide Distribution including USA and countries of Australia, Belgium. Canada, Columbia, Dominican Republic, Egypt, Germany, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Ma
- Date initiated
- 2008-07-18
- Date posted
- 2008-12-11
- Date terminated
- 2009-07-07
- Location
- Ann Arbor, MI