Terumo Advanced Perfusion System 1 (4 inch diameter Roller Pump, Gray; Model 816570). Extracorporeal circulation of bloo

FDA device recall Z-0356-2009 · posted 2008-12-11 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corporation
Reason for recall
Overspeed and underspeed errors, erratic pump behavior, jerky operation at low RPMs, pump instability, pump slowdowns and pump stops due to overspeed max events may occur due to a grease leak onto the motor speed encoder disk.
Action taken
Consignees were notified of the problem via recall letter dated 7/18/08. The firm began conducting field upgrades on the roller pumps in 07/08. For questions or concerns, contact Terumo CVS Customer Service at 1-800-521-2818.
Root cause
Component design/selection
Status
Terminated
Product code
DTQ
Quantity affected
269
Distribution
Worldwide Distribution including USA and countries of Australia, Belgium. Canada, Columbia, Dominican Republic, Egypt, Germany, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Ma
Date initiated
2008-07-18
Date posted
2008-12-11
Date terminated
2009-07-07
Location
Ann Arbor, MI