Terumo Advanced Perfusion System 1, 4 inch diameter Roller Pump; Model 801040.
FDA device recall Z-0438-2007 · posted 2007-02-14 · Terminated
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corp
- Reason for recall
- The roller pump will pause in response to a pressure alarm condition, as designed, but then may not resume operation as expected after the alarm condition has been cleared.
- Action taken
- Consignees were notified of the problem via letter dated 11/10/06 and informed that they would be contacted to schedule a service call to install software upgrades. Installation of software upgrades began at customer locations on 10/16/06.
- Root cause
- Other
- Status
- Terminated
- Product code
- DTQ
- Quantity affected
- 1,399
- Distribution
- Worldwide distribution ---- including USA and countries of Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China
- Date initiated
- 2006-10-16
- Date posted
- 2007-02-14
- Date terminated
- 2009-07-07
- Location
- Ann Arbor, MI