Terumo Advanced Perfusion System 1, 4 inch diameter Roller Pump; Model 801040.

FDA device recall Z-0438-2007 · posted 2007-02-14 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corp
Reason for recall
The roller pump will pause in response to a pressure alarm condition, as designed, but then may not resume operation as expected after the alarm condition has been cleared.
Action taken
Consignees were notified of the problem via letter dated 11/10/06 and informed that they would be contacted to schedule a service call to install software upgrades. Installation of software upgrades began at customer locations on 10/16/06.
Root cause
Other
Status
Terminated
Product code
DTQ
Quantity affected
1,399
Distribution
Worldwide distribution ---- including USA and countries of Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China
Date initiated
2006-10-16
Date posted
2007-02-14
Date terminated
2009-07-07
Location
Ann Arbor, MI