Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base, Terumo
FDA device recall Z-0910-2009 · posted 2009-02-24 · Terminated
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Reason for recall
- The power manager board may cause a short circuit, resulting in loss of system power, loss of central control monitor or electrical arcing.
- Action taken
- An "Urgent Medical Device Correction" letter was sent to consignees on November 7, 2008. The letter described the reason for correction and provided instructions to customers. Terumo Cardiovascular Systems Corporation will replace the power manager boards on the affected devices. Customers were instructed to complete and fax the Customer Response Form to CVS Customer Service at 1-800-292-6551 (US only) or respond by telephone at 1-800-521-2818. Please direct questions to CVS Customers Service at 1-800-521-2818.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- DTQ
- Quantity affected
- 515
- Distribution
- Worldwide Distribution including USA and countries of Australia, Belgium, Canada, Columbia, Costa Rica, Dominican Republic, Egypt, Germany, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan
- Date initiated
- 2008-11-07
- Date posted
- 2009-02-24
- Date terminated
- 2009-07-07
- Location
- Ann Arbor, MI