Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base; Model 8

FDA device recall Z-0930-2007 · posted 2007-06-30 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corp
Reason for recall
System 1 power supply may electrically short to the main console, disrupting normal system operation by causing unexpected re-setting or shutdown of pumps and modules, false alarms and/or overheating and failure of the central control monitor.
Action taken
The firm corrected the complaint units at the time of the complaints in 2005. The firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and advising them that their system will be corrected when replacement components become available.
Root cause
Other
Status
Terminated
Product code
DTQ
Quantity affected
416 for both models
Distribution
Worldwide: USA, Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, Pakistan and Singapore.
Date initiated
2005-02-01
Date posted
2007-06-30
Date terminated
2009-07-07
Location
Ann Arbor, MI