Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base; Model 8
FDA device recall Z-0603-2007 · posted 2007-06-30 · Terminated
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corp
- Reason for recall
- The power supply may fail to function due to a defective capacitor. (Note; this device incorporates a redundant separate power supply not affected by this capacitor)
- Action taken
- The power supplies were replaced as complaints were received between August 2005 and February 2006. The firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and advising them that their units will be corrected when replacement components become available.
- Root cause
- Other
- Status
- Terminated
- Product code
- DTQ
- Quantity affected
- 534 of both model numbers
- Distribution
- Worldwide Distribution: USA, Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Jordan, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia,
- Date initiated
- 2005-08-01
- Date posted
- 2007-06-30
- Date terminated
- 2009-08-04
- Location
- Ann Arbor, MI