Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required); Model 801764

FDA device recall Z-0209-05 · posted 2004-11-18 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corp
Reason for recall
The central control monitor screen may go blank during use because of a bad wiring connection in the wiring harness or a low voltage setting on the potentiometer.
Action taken
Service representatives were sent a service bulletin dated 8/12/03 instructing them to correct units in the field bearing serial numbers between 20 and 109.
Root cause
Other
Status
Terminated
Product code
DTQ
Distribution
United States, Belgium, Canada, Dubai, Egypt, Germany, Hong Kong, India, Japan, Korea, Singapore and United Kingdom.
Date initiated
2003-08-12
Date posted
2004-11-18
Date terminated
2005-09-22
Location
Ann Arbor, MI