Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required); Model 801764
FDA device recall Z-0209-05 · posted 2004-11-18 · Terminated
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corp
- Reason for recall
- The central control monitor screen may go blank during use because of a bad wiring connection in the wiring harness or a low voltage setting on the potentiometer.
- Action taken
- Service representatives were sent a service bulletin dated 8/12/03 instructing them to correct units in the field bearing serial numbers between 20 and 109.
- Root cause
- Other
- Status
- Terminated
- Product code
- DTQ
- Distribution
- United States, Belgium, Canada, Dubai, Egypt, Germany, Hong Kong, India, Japan, Korea, Singapore and United Kingdom.
- Date initiated
- 2003-08-12
- Date posted
- 2004-11-18
- Date terminated
- 2005-09-22
- Location
- Ann Arbor, MI