Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required) base; Model

FDA device recall Z-0602-2007 · posted 2007-06-30 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corp
Reason for recall
The power supply may fail to function due to a defective capacitor. (Note; this device incorporates a redundant separate power supply not affected by this capacitor)
Action taken
The power supplies were replaced as complaints were received between August 2005 and February 2006. The firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and advising them that their units will be corrected when replacement components become available.
Root cause
Other
Status
Terminated
Product code
DTQ
Quantity affected
534 of both model numbers
Distribution
Worldwide Distribution: USA, Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Jordan, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia,
Date initiated
2005-08-01
Date posted
2007-06-30
Date terminated
2009-08-04
Location
Ann Arbor, MI