Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required) base; Model
FDA device recall Z-0432-2007 · posted 2007-02-14 · Terminated
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corp
- Reason for recall
- A 'battery needs service' message may be displayed, or the remaining battery time may incorrectly be reported out as zero, even when the battery is fully functional.
- Action taken
- Consignees were sent a letter dated 11/10/06 informing them of the nature of the issues and that the firm would contact them to schedule a service call to install the software upgrade package. The software upgrade installation began at customers on 10/16/06, prior to issuance of the letter to consignees.
- Root cause
- Other
- Status
- Terminated
- Product code
- DTQ
- Quantity affected
- 489
- Distribution
- Worldwide distribution ---- including USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, S
- Date initiated
- 2006-10-16
- Date posted
- 2007-02-14
- Date terminated
- 2009-07-07
- Location
- Ann Arbor, MI