Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required) base; Model

FDA device recall Z-0598-2007 · posted 2007-06-30 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corp
Reason for recall
The power manager subsystem may not initialize properly during start-up, which will cause the battery status monitoring and battery switching operations to not be fully functional.
Action taken
Firm repaired the affected units beginning in November 2005, upon receipt of complaints of failure. The firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and advising them that they will be visited and their units upgraded as replacement components become available.
Root cause
Other
Status
Terminated
Product code
DTQ
Quantity affected
485 of both model numbers
Distribution
Worldwide distribution ---- Nationwide and countries of Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Philippines, Rep
Date initiated
2005-11-01
Date posted
2007-06-30
Date terminated
2009-06-23
Location
Ann Arbor, MI