Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required); Models 801
FDA device recall Z-1286-06 · posted 2006-07-27 · Terminated
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corp
- Reason for recall
- While on battery power, the system may flash a low battery warning even though the battery is adequately charged.
- Action taken
- Customers were notified via letter dated October 19, 2005 that the Power Manager Software would be upgraded within the next six months. As of 7/7/06 all units within the U.S. have been completed and international units are still in the process of being upgraded.
- Root cause
- Other
- Status
- Terminated
- Product code
- DTQ
- Quantity affected
- 329
- Distribution
- Nationwide and worldwide, including Australia, Belgium, Canada, China, Dominican Republic, Egypt, Georgia, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, S
- Date initiated
- 2005-10-19
- Date posted
- 2006-07-27
- Date terminated
- 2008-06-19
- Location
- Ann Arbor, MI