Terumo Advanced Perfusion System 1: 100/120V AC, Advanced Perfusion System 1 The Terumo Advanced Perfusion System 1 is i

FDA device recall Z-1968-2013 · posted 2013-08-14 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corporation
Reason for recall
The operators manual does not fully describe the sequence of expected events that will occur when a specific operating condition exists. Specifically, if the user is adjusting the System 1 centrifugal pump speed via the Central Control Monitor at the same time an alarm condition occurs with configuration for a pump coast response, the labeling in the operators manual does not fully describe how the pump will react.
Action taken
Terumo CVS, sent an URGENT MEDICAL DEVICE RECALL notification letter dated July 26, 2013, to all affected customers explaining the correction to the operators manual, potential hazards and what to do in response to the issue. If you have any questions or concern please contact Terumo CVS Customer Service 1-800-521-2818 Recall Fax 1-734-741-6149 Customer Service Hours: Monday Friday, 8 AM - 6 PM ET
Root cause
Software design
Status
Terminated
Product code
DTQ
Quantity affected
1657
Distribution
Worldwide Distribution - USA (nationwide) including the countries of Australia, Dubai, Hong Kong, Indonesia, Singapore, China, Mexico, British Columbia, Taiwan, Thailand, Chile, Malaysia, Japan, Belgi
Date initiated
2013-07-26
Date posted
2013-08-14
Date terminated
2017-02-07
Location
Ann Arbor, MI