Teleflex Medical, Research Triangle Park, NC, Polypropylene, Rx Only, Sterile, Distributed by Boston Scientific, Natick,

FDA device recall Z-1557-2014 · posted 2014-05-07 · Terminated

Recall details

Recalling firm
Teleflex Medical
Reason for recall
Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached from the suture during use.
Action taken
Urgent Medical Device Recall Notification letters were issued on March 11, 2014 to consignees. The letter requested that customers discontinue use and quarantine any product on hand. The attached Recall Acknowledgment Form should be completed and returned to Teleflex regardless of whether affected product is on hand or not. Once received a customer service representative will contact customers with instructions on how to return the affected product. Questions should be directed to Customer Service at 1-866-246-6990.
Root cause
Employee error
Status
Terminated
Product code
GAQ
Quantity affected
32,271 ea.
Distribution
Worldwide Distribution -- USA, including the states of AL, CA, FL, GA, IL, IN, KS, LA, ME, MA, MI, MN, MS, MO, NV, NJ, NM, NY, OH, PA, RI, SC, TN, TX, UT, VA, WV, and WI; and countries of Canada, Germ
Date initiated
2014-03-11
Date posted
2014-05-07
Date terminated
2017-08-14
Location
Durham, NC