Tecan Infinite M1000 The Tecan Infinite M1000 is a multifunctional monochromator based microplate reader that provides h

FDA device recall Z-2664-2016 · posted 2016-08-25 · Terminated

Recall details

Recalling firm
Tecan US, Inc.
Reason for recall
The batteries have the potential to overheating, melt or char.
Action taken
Tecan contacted all customers by Safety Announcement, via certified mail starting July 25, 2016. The letter identified the product, the problem, and the action to be taken by the customer. Because affected batteries pose a fire and burn hazard, it is extremely important that customers check whether their battery is affected. Customers were advised to go immediately to the HP Battery Recall website at: http://www.HP.com/go/batteryprogram2016 to check if their battery is affected by this recall. Customers were also advised to complete and return the Customer Response Form at QA@tecan.com or fax it to 919-287-2961. Customers with questions should contact HP at 1-888-202-4320 or Tecan Technical Service's Helpdesk at 800-832-2687, option 1.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
JJQ
Quantity affected
1 unit
Distribution
Worldwide Distribution - US including AK, AL, AR, CA, FL, GA, HI, KY, MA, MD, NC, NY, PA, PR, SC, TN, TX, UT, VA. International account: Canada.
Date initiated
2016-07-25
Date posted
2016-08-25
Date terminated
2017-01-23
Location
Morrisville, NC